Philips MP20/30 Medical Alarms User Manual


 
Technical Alarm Messages (INOPs) 4 Patient Alarms and INOPs
69
NBP INTERRUPTED
Numeric is replaced by -?-
INOP tone.
NBP Check the tubing and cuff for leakages or kinks. Check that
you are using the correct cuff size and placement, and that the
correct patient category is selected. Try restarting the
measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the Stop All
SmartKey is selected.
This INOP arises when the measurement needed longer than
the maximum time for inflation, deflation or the total
measurement.
NBP MEASURE FAILED
Numeric may be displayed with a -?-
INOP tone.
NBP Check that you are using the correct cuff size and placement,
and that the correct patient category is selected. Try restarting
the measurement.
If numerics are displayed, the monitor is able to measure mean
only and the alarm source is set to S, D or S&D.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the Stop All
SmartKey is selected.
Check the condition and suitability of the patient for NBP
monitoring. Use another cuff to continue measuring.
No Central Monit.
INOP tone
Monitor There is a problem with the communication to the network.
Central monitoring is currently not possible (no patient alarms
or information). Check the connection. Contact your service
personnel.
NO PPV FROM MeasSrv
at Information Center
MMS or
FMS
The measurement server does not supply a beat-to-beat arterial
pressure value. Contact your service personnel.
NO PPV FROM <Server> MMS or
FMS
The measurement server does not supply a beat-to-beat arterial
pressure value. Contact your service personnel.
P/P1/P2/P3/P4 INOPS PRESS See <Pressure label> INOPS (under Pressure).
PAP INOPS PRESS See <Pressure label> INOPS (under Pressure).
PPV BAD <Pressure Label>
SIGNAL
PRESS The arterial pressure source selected for PPV is not providing a
pulsatile signal.
PPV BAD SIGNAL
at Information Center
PRESS The arterial pressure source selected for PPV is not providing a
pulsatile signal.
PPV CHK SOURCES PRESS The arterial pressure source selected for PPV is unplugged or
switched off. When this INOP has displayed for 1 minute PPV
will be switched off.
<Pressure label> ARTIFACT
Numeric displayed with -?-
PRESS A non-physiological event is detected (for example, a flush or
blood sample). A resulting high limit alarm will be suppressed.
<Pressure label>
DEACTIVATED
INOP tone
PRESS A Pressure measurement label in the measurement server or
extension has been deactivated, either by connecting a Temp
transducer in the shared Press/Temp socket, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a
Pressure transducer or reactivate the measurement label in the
Measurement Selection window.
INOP Message, Indication Source What to do