Philips MP20/30 Medical Alarms User Manual


 
34 Installation and Specifications Measurement Specifications
360
Volume Wave (AWV) Range 0 to 100 ml (neonatal
0 to 400 ml (pediatric)
0 to 3000 ml (adult)
Accuracy
±5% or
±3 ml (neonatal)
±12 ml (pediatric)
±90 ml (adult),
whichever is greater
Scales
50, 100, 200, 600, 800, 1000, 2000, 3000 ml
Scale Default
50 ml (neonatal
200 ml (pediatric)
800 ml (adult)
Pressure Wave (AWP) Range -20 to 120 cmH
2
O
Accuracy
±2% or ±0.5 cmH2O whichever is greater
Scales
10, 20, 40, 60, 80, 100, 120 cmH
2
O
Scale Default
20 l/min (neonatal
40 l/min (pediatric)
40 l/min (adult)
Respiration Rate
numeric (RRspir)
Range 2 to 120 breaths/min
Resolution
1 breath/min
Inspired Minute
Volume numeric
(MVin)
Range
0.01 to 5 l/min (neonatal)
0.06 to 30 l/min (pediatric)
0.4 to 60l/min (adult)
Resolution
0.01 l/min (neonatal/pediatric)
0.1 l/min (adult)
Expired Minute
Volume numeric
(MVexp)
Range
0.01 to 5 l/min (neonatal)
0.06 to 30 l/min (pediatric)
0.4 to 60l/min (adult)
Resolution
0.01 l/min (neonatal/pediatric)
0.1 l/min (adult)
Inspired Volume
numeric (TVin)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution
1 ml (neonatal/pediatric))
10 ml (adult)
Expired Volume
numeric (TVexp)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution
1 ml (neonatal/pediatric)
10 ml (adult)
Lung Compliance
numeric
Range 0 to 500 ml/cmH
2
O
Resolution
1 ml/cmH
2
O
Airway Resistance
numeric
Range 0 to 500 cmH
2
O/l/s (neonatal)
0 to 250 cmH
2
O/l/s (pediatric)
0 to 100 cmH
2
O/l/s (adult)
Resolution
1 cmH
2
O/l/s
Peak Inspiratory
Pressure numeric
(PIP)
Range 0 to +120 cmH
2
O
Resolution
1 cmH
2
O
Spirometry Performance Specifications