34 Installation and Specifications Monitor Performance Specifications
344
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data
configurable (except MP20/30), either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
For MP20/30: 4 minutes of high resolution trend
Capacity (max)
25 or 50 events for either 8 or 24 hours
Alarm signal System delay less than 3 seconds
Pause duration
1,2,3 minutes or infinite, depending on configuration
Extended alarm pause
5 or 10 minutes
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity
100 items
Real Time Clock Range from: January 1, 1997, 00:00 to: December 31, 2080, 23:59
Accuracy
< 2 seconds per day (typically)
Hold Time
infinite if powered by AC; otherwise at least 48 hours (typical:
> 72 hours)
Buffered Memory Hold Time if powered by AC: infinite
without power: at least 48 hours (typical: > 72 hours)
Contents
Active settings, trends, patient data, realtime reports, events,
review alarms
MMS M3001A Performance Specifications
Tre n d s Trend Data Buffered
Memory
if powered by monitor mains connection via MSL: infinite
without power applied: at least 6 hours
Contents
trend data, patient identification and all active settings
D80 Intelligent Display Performance Specifications
Power Specifications Power consumption < 145 W
Line Voltage 100 to 240 V ~
Current 1.6 to 0.7 A
Frequency 50/60 Hz ~
Docking Station M8045A Performance Specifications
Power Line Voltage 100 - 240V
Frequency
50/60 Hz
Power Consumption
negligible (for Power On LED)
Current to attached
patient monitor
max. 3.5 A
MP20/30, MP40/50, MP60/70/80/90