Philips MP20/30 Medical Alarms User Manual


 
4 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
68
LEADS OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
ECG Check that all of the required ECG leads are attached, and that
none of the electrodes have been displaced.
LL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
ECG The LL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
New
Lead Setup
in the Setup ECG menu to confirm the
new lead set.
MCC Reversed Monitor The MSL coupling cable is reversed. Connect the end with the
grey connector to the Intelligent Display.
MCC Unsupported Monitor An MSL coupling cable has been connected to a device which
does not support MSL coupling.
Measserv Unsupportd Monitor The measurement server is not supported by the monitor.
Contact your service personnel.
MMS UNPLUGGED
INOP tone.
MMS Make sure that the Multi-Measurement Server is connected to
the monitor. All MMS measurements are off while the MMS is
unplugged.
MMS UNSUPPORTED
INOP tone.
MMS The Multi-measurement Server is not supported by your
monitor. Contact your service personnel.
MSL Power High Monitor The power consumption of the devices connected to the
Measurement Server Link (MSL) cable is too high. If this
situation continues, the MSL will be switched off. Contact
your service personnel.
MSL Power Off
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Server Link (MSL) cable was too high for too
long and the MSL has been switched off. Contact your service
personnel.
MSL Power Overload
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Server Link (MSL) cable is much too high or
there has been a short circuit. The MSL has been switched off.
Contact your service personnel.
NBP CUFF OVERPRESS
Numeric displayed with -?- ; INOP
tone.
During this INOP, alarms cannot be
paused or switched off.
NBP The NBP cuff pressure exceeds the overpressure safety limits.
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the Stop All
SmartKey is selected.
NBP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
NBP Remove the cuff from the patient. The NBP hardware is faulty.
Contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the Stop All
SmartKey is selected.
INOP Message, Indication Source What to do