Philips MP20/30 Medical Alarms User Manual


 
Measurement Specifications 34 Installation and Specifications
361
Cardiac Output / Continuous Cardiac Output
Negative Inspiratory
Pressure numeric
Range -20 to 0 cmH
2
O
Resolution
1 cmH
2
O
Mean Airway Pressure Range -20 to +120 cmH
2
O
Resolution
1 cmH
2
O
Plateau Pressure
numeric
Range 0 to +120 cmH
2
O
Resolution
1 cmH
2
O
Positive End-
Expiratory Pressure
numeric (PEEP)
Range 0 to 99 cmH
2
O
Resolution
1 cmH
2
O
Barometric Pressure
numeric (PB)
Range 400 to 800 mmHg
Resolution
1 mmHg
Peak Inspiratory Flow
numeric (PIF)
Range 0.1 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution
0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Peak Expiratory Flow
numeric (PEF)
Range 0.1 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution
0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Inspiration to
Expiration ratio
numeric (I:E)
Range 1: 0.2 ... 9.9
Resolution
0.1
Spirometry Performance Specifications
C.O./CCO Performance Specifications
C.O. (right heart) Range 0.1 to 20.0 l/min
Accuracy
Instrument Specification (electrical):+ 3% or 0.1 l/min
System Specification:+ 5%.or 0.2 l/min
Repeatability
Instrument Specification (electrical):+ 2% or 0.1 l/min
System Specification:+ 3%.or 0.1 l/min
C.O.
(transpulmonary)
Range 0.1 to 25.0 l/min
Accuracy
Instrument Specification (electrical):+ 4% or 0.15 l/min
System Specification:+
5% or 0.2 l/min
Repeatability
Instrument Specification (electrical):+ 2% or 0.1 l/min
System Specification:+
3% or 0.1 l/min
EVLW
not available in USA
Range 10 to 5000 ml
Standard Deviation
10% or 1 ml/kg