NBP Default Settings 35 Default Settings Appendix
375
NBP Default Settings
Temperature Default Settings
Pulse Alarms On/Off on on on
Pulse High Limit 120 bpm 160bpm 200 bpm
Pulse Low Limit 50 bpm 75 bpm 100 bpm
∆ Extr Brady 20 bpm 20 bpm 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm
∆ ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
SpO
2
Alarm Default Settings
Setting adult pediatric neonatal
NBP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Auto/Manual Auto Manual
Alarms from Sys.
High Alarm Limit 160/ 90 (110) 120/ 70 (90) 90/ 60 (70) 180/ 90 (110)
Low Alarm Limit 90/ 50 (60) 70/ 40 (50) 40/ 20 (24) 70/ 50 (65)
Alarms On
NBP On
Repetition Time 15 min 3 min
Unit mmHg
Done Tone Off On
Start Time Synchronized NotSynchron.
VP Pressure 60 mmHg 40 mmHg 30 mmHg
Reference Auscultatory Invasive
Color Red Magenta
Temp Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Low Limit 36 35
High Limit 39
Alarms On
Unit
0
C
Range 35...43 30 ... 43
Color Green Light Green