Philips MP20/30 Medical Alarms User Manual


 
Patient Alarm Messages 4 Patient Alarms and INOPs
55
**HR HIGH ECG The heart rate high alarm limit has been
exceeded.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone. The sound
switches off after 5 seconds if
Arrhythmia is On.
**HR LOW ECG The heart rate has fallen below the low
alarm limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone. The sound
switches off after 5 seconds if
Arrhythmia is On.
**imCO2 HIGH CO
2
, Resp,
AGM
The inspired minimum CO
2
high alarm
limit has been exceeded.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**in <Agent label>
HIGH
AGM The inspired agent high alarm limit has
been exceeded.
numeric flashes, high limit is
highlighted, yellow alarm
lamp, alarm tone.
**in <Agent label>
LOW
AGM The inspired agent value has fallen
below the AGT low alarm limit.
numeric flashes, low limit is
highlighted, yellow alarm
lamp, alarm tone.
**inN2O HIGH N
2
O, AGM The inspired N
2
O high alarm limit has
been exceeded.
numeric flashes, high limit is
highlighted, yellow alarm
lamp, alarm tone.
**inO2 HIGH O
2
, AGM The inspired O
2
high alarm limit has
been exceeded.
numeric flashes, high limit is
highlighted, yellow alarm
lamp, alarm tone.
**inO2 LOW O
2
, AGM The inspired O
2
value has fallen below
the low alarm limit.
numeric flashes, low limit is
highlighted, yellow alarm
lamp, alarm tone.
***inO2 LOW OXYGEN O
2
, AGM The inspired O
2
value has fallen below
18 vol.%.
numeric flashes, low limit is
highlighted, red alarm lamp,
alarm tone.
*/**IRREGULAR HR ECG/
Arrhythmia
Consistently irregular heart rhythm. numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**MISSED BEAT ECG/
Arrhythmia
No beat detected for 1.75*R-R interval,
or if HR>120bpm no beat detected for
one second (non-paced patients only).
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
*/**MULTIFORM PVCs ECG/
Arrhythmia
Two differently shaped Vs detected,
each occurring at least twice within the
last 300 beats and at least once within
the last 60 beats.
numeric flashes, yellow alarm
lamp, short yellow audible
alarm.
**NBP HIGH NBP The measured NBP value is above the
high alarm limit.
s, d, or m after the label indicates
whether the systolic, diastolic or mean
pressure has crossed the limit.
numeric flashes and high limit
is highlighted, yellow alarm
lamp, alarm tone.
**NBP LOW NBP The measured NBP value is below the
low alarm limit.
s, d, or m after the label indicates
whether the systolic, diastolic or mean
pressure has crossed the limit.
numeric flashes and low limit
is highlighted, yellow alarm
lamp, alarm tone.
Alarm Message From Condition Indication