Leica DM1000 LED Microscope & Magnifier User Manual


 
44
9. Contrast Methods
9.1.3 Darkfield*
Insert a transmitted light specimen
Rotate an appropriate objective into place
Bring the image into focus using the focus dial
and set the brightness
Condenser UCL/P:
Click the condenser disk into the BF position
Condensers CL/PH, CLP/PH, and
APL .ACHR.0.9 (P):
Pull out the DF light ring slide as far as the
stop;
check the Köhler illumination (
p. 28)
Open the aperture diaphragm completely
(position PH)
Condenser UCL/P:
Click the condenser disk into the DF position
Condensers CL/PH, CLP/PH, and
APL.ACHR.0.9 (P):
Insert the DF light ring slide as far as the stop
Note:
Condensers UCL/UCLP and UCA/P: The DF light
ring must be centered (
p. 29).
Special dark field condensers* are available for
the DM1000/DM1000 LED.
The application potential of the DF condensers
depends on the aperture of the objective in use.
For objectives with built-in iris diaphragm, the
aperture can be adapted.
DF condenser max. objective aperture
D 0.80 - 0.95 0.75
D 1.20 - 1.44 OIL 1.10
9.1.2 Phase Contrast*
Insert a transmitted light specimen
Rotate an appropriate objective into place;
objectives that are suitable for phase contrast
are engraved with PH
Bring the image into focus using the focus dial
and set the brightness
For an optimal field diaphragm setting, check
the Köhler illumination (
p. 28)
Open the aperture diaphragm completely
(position PH)
Condensers UCL/UCLP and UCA/P:
Set the light ring corresponding to the
objective on the condenser disk
Example: Light ring 1 belongs to the objective
with the engraving PH 1
Condensers CL/PH, CLP/PH, and
APL. ACHR.0.9 (P):
Use the light ring slide
Note:
Condensers UCL/UCLP and UCA/P: Light rings
must be centered (
p. 29).