7. Start-up
Caution:
In older lamps, the structure of the arc is no
longer clearly recognizable. The image is
then more like that of a Hg 50 lamp. The im-
age and mirror image can no longer be su-
perimposed exactly. In this case, align both
images.
Using the collector, defocus the image with
the knob (30.6) until the arc image and mirror
image are no longer recognizable and the im-
age is homogeneously illuminated