Leica DM1000 LED Microscope & Magnifier User Manual


 
38
8. Operation
HC L 2TU tube:
The beam splitting is set manually by pulling out
a control bar.
Control Bar Observation Photo
VIS 100% 0%
PHOTO 110% 100%
8.5 Eyepieces
Note:
The eyepiece’s aperture protector must be
removed or folded back during microscopy
while wearing eyeglasses.
We recommend
that users remove their glasses
with bifocal or progressive-addition lenses
when working with the microscope.
For the adjustable tubes with documentation
output, choose the 100% VIS position.
Eyepieces with Inlaid Reticle*
Focus the reticle by adjusting the eyelens
Focus on the object through this eyepiece
Then, close that eye and focus on the object
by adjusting only the second ocular
Correction for Vision Problems
With your right eye, look through the right
eyepiece and bring the specimen into sharp
focus
Then, with your left eye, view the same
specimen and rotate the left eyepiece tube
until the object is brought into sharp focus;
do not use the focus dial
Fig. 42 Tube range HC L (extract)
1 Binocular observation tube HC LB 0/3/4
2 Ergonomy tube HC LVB 0/4/4, binocular,
viewing angle 0-35°
additional ergotube (short) HC -/0/4,
swivelable 7.5°-32.5°
3 Trinocular tube H L1T 4/5/7, with fixed beamsplitter
(50%/50%)
4 HC L1VT 0/4/4 like 3,
but with adjustable viewing angle of 0-35°
5 Photo adapter, with 2 exits (50%/50%)
6 Photo TV exit
2
34
56
1