Leica DM1000 LED Microscope & Magnifier User Manual


 
21
6. Assembly
6.6 Components for Fluorescence Applications*
6.6.1 Fluorescence illuminator*
The fluorescence illuminator is mounted before
the tube. It is fastened in place with the side
clamping screw (13.1).
6.6.2 106z Lamp Housing*
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
During assembly, always unplug the power
supply unit of the 106z lamp housing from its
socket.
During assembly work on xenon burners, al-
ways wear the supplied protective gloves
and face protection (fig. 14) (risk of explosion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
The lamp housing 106z is used with various gas
discharge lamps.
Fig. 14
Protective gloves and mask
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow at least 30 mins
for cooling down!
Inserting the Gas Discharge Lamps* (Hg and
Xe) into the 106z Lamp Housing
Hg and Xe lamps are powered by separate
supply units.
Read the separate instruction manual provided
with these supply units.
Fig. 13 Assembly of fluorescence illuminator
1 Clamping screw
1