7. Start-up
7. Start-up
7.1 Switching on the Microscope
• Switch on the microscope with the on/off
switch (24.1)
Caution!
After turning on the gas discharge lamp*,
the burner must be immediately adjusted.
Therefore, do not turn on the power supply*
unit yet. First, work in transmitted light in or-
der to familiarize yourself with the micro-
scope’s controls.
7.2 Köhler Illumination*
The condenser is also pre-adjusted in the
factory.
However, it may be necessary to re-adjust the
condenser in some cases. Therefore, check the
condenser centering.
The following procedure is provided for the
transmitted light brightfield illumination.
If present click the condenser disk* into the
BF position
If present pull the light ring slide* out of the
condenser
Select an objective with moderate
magnification (10x-20x);
for condensers with movable condenser
heads:
Swing in the condenser top
(The condenser top is swung out for objective
magnifications < 10x.)
Insert the specimen into the stage’s specimen
holder
Focus on the specimen using the focus wheel
(24.2)
Set the light intensity using the brightness
control (25.2)
Close the field diaphragm (25.3) until the edge
of the diaphragm appears in the specimen
plane (26a)