Invacare 1080556 Wheelchair User Manual


 
Part No. 1080556 Rev. I 27 Solara™
This Procedure Includes the Following:
Installing/Removing/Adjusting Dual Point Armrest
Using/Installing/Adjusting Locking Cantilever Arms
Using/Installing/Adjusting Non-Locking Cantilever Arms
Arm Pad Replacement for Cantilever Arms
Installing/Removing T-Arms
Adjusting the T-Arms
Adjusting Transfer Assists and/or Side Guards - Arms
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attach-
ing hardware is tightened securely - otherwise injury or damage may occur.
INSTALLING/REMOVING/ADJUSTING DUAL POINT
ARMREST (FIXED AND ADJUSTABLE HEIGHT) (FIGURE 1)
WARNING
Make sure the locking mechanism is secured BEFORE using the wheelchair.
INSTALLING ARMREST
1. Unlock armrest assembly by flipping the arm assembly release lever located on the side rail
to the UP (HORIZONTAL) position as shown in DETAIL "A" .
NOTE: Armrest locks MUST be in the unlocked position when placing arm into the arm sockets.
2. Install armrest assembly into the arm sockets.
3. When installed, lock the armrest assembly by pressing the release lever into the DOWN
(VERTICAL) position as shown in DETAIL "A".
4. Repeat STEPS 1-3 for the opposite side of the wheelchair.
5. If necessary, adjust the armrest height. Refer to
ADJUSTING ARMREST in this procedure
of the manual.
REMOVING ARMREST
1. Unlock armrest assembly by flipping the arm assembly release lever located on the side rail
to the UP (HORIZONTAL) position as shown in DETAIL "A".
2. Pull up on the armrest assembly to remove the armrest from the arm socket.
3. Repeat STEPS 1-2 for opposite side of wheelchair, if necessary.
ARMS
ARMS
PROCEDURE 5ARMS