2-19
Rosemount 2051
Inline Process Connection
Housing Rotation The electronics housing can be rotated up to 180 degrees in either direction to
improve field access, or to better view the optional LCD display. To rotate the
housing, perform the following procedure:
1. Loosen the housing rotation set screw using a -in. hex wrench.
2. Turn the housing left or right up to 180° from its original position. Over
rotating will damage the transmitter.
3. Retighten the housing rotation set screw.
Figure 2-10. Housing Rotation
Do not apply torque directly to the sensor module. Rotation between the sensor module
and the process connection can damage the electronics. To avoid damage, apply torque
only to the hex-shaped process connection.
Sensor Module
Process Connection
5
64
------
Housing Rotation Set
Screw (5/64-in.)