Emerson Process Management 2051 Sleep Apnea Machine User Manual


 
Reference Manual
00809-0200-4101, Rev AA
July 2008
Rosemount 2051
C-10
Simulate_STATE 66 The state of the simulate jumper.
0 = Uninitialized
1 = Jumper/ switch off, simulation not allowed
2 = Jumper/ switch on, simulation not allowed (need to cycle jumper/ switch)
simulate_IO 64 Status of simulate jumper/switch.
RB_SFTWR_REV_ALL 51 The string will contain the following fields:
Major rev: 1-3 characters, decimal number 0-255
Minor rev: 1-3 characters, decimal number 0-255
Build rev: 1-5 characters, decimal number 0-255
Time of build: 8 characters, xx:xx:xx, military time
Day of week of build: 3 characters, Sun, Mon,...
Month of build: 3 characters, Jan, Feb...
Day of month of build: 1-2 characters, decimal number 1-31
Year of build: 4 characters, decimal
Builder: 7 characters, login name of builder
RB_SFTWR_REV_BUILD 50 Build of software that the resource block was created with.
RB_SFTWR_REV_MAJOR 48 Major revision of software that the resource block was created with.
RB_SFTWR_REV_MINIOR 49 Minor revision of software that the resource block was created with.
start_with_defaults 63 0 = Uninitialized
1 = do not power-up with NV defaults
2 = power-up with default node address
3 = power-up with default pd_tag and node address
4 = power-up with default data for the entire communications stack (no application data)
STRATEGY 03 The strategy field can be used to identify grouping of blocks.
ST_REV 01 The revision level of the static data associated with the function block.
summary_status 56 An enumerated value of repair analysis.
TAG_DESC 02 The user description of the intended application of the block.
TEST_RW 08 Read/write test parameter - used only for conformance testing.
UPDATE_EVT 35 This alert is generated by any change to the static data.
WRITE_ALM 40 This alert is generated if the write lock parameter is cleared.
WRITE_LOCK 34 If set, no writes from anywhere are allowed, except to clear WRITE_LOCK. Block inputs
will continue to be updated.
WRITE_PRI 39 Priority of the alarm generated by clearing the write lock.
Parameter
Index
Number
Description