Barco K5902007 - 00 Sleep Apnea Machine User Manual


 
Using MediCal Pro 68
3.3 Changing passwords
When you have logged in as advanced user, you can change the
advanced user password.
To change the advanced user password:
1 From the main window, press Alt-A on the keyboard or select Tools
Options User Login... from the menus. Your user name is still filled
out in the User name field.
2Click on Change Password.
3 In field Old password, fill in the current password.
4 In fields New password and Re-enter new password, fill in the new
password.
5Click on OK. The password is changed.
6 The Login dialog appears again. Click on Cancel to continue as
advanced user, or log in as standard user.
3.4 Calibrating displays
The calibration of the displays is a very important issue. Calibration is
necessary to obtain conformance and to maintain consistency. It is
inevitable during Configuration setup and may be necessary as
intervention when a Q/A check turns out to be Not OK.
You can calibrate Barco displays and non-Barco displays, but the
procedure for both categories is different.
MediCal Pro supports the following sensors for calibration: Barco LCD
sensor, Barco Serial Sensor, X-Rite DTP92 (serial and USB), Wellhofer Luxor
LXPlus, Solar PMA2200, Minolta LS100, Verilum CPOD (serial and USB) and
Tektronix TJ17. Also the BARCO I-Guard sensor inside BARCO flat panel
displays is supported.
Important:
The displays must be warmed up for at least 5 minutes before starting
calibration.
To start calibration:
Calibration starts automatically after defining a new, uncalibrated
Preset or after modifying a Preset.