Barco K5902007 - 00 Sleep Apnea Machine User Manual


 
First installation 38
Display function
Select the required display function from the drop-down list. For most
medical viewing applications, we recommend you select the DICOM
display function, as this function results in the most visible grayscales.
If you select Gamma, you can enter a display gamma value if the imaging
board head, to which the display is connected, has an editable gamma
table.
If you select Uncorrected, the program fills the imaging board gamma
table with a linear function.
Ambient light compensation
Check this option if you want to compensate for the influence of the
ambient light on the image. Especially when you have selected a DICOM
or CIE display function, we advise to switch on this compensation to
obtain a better display performance.
Glare compensation
Check Glare Compensation if you want to compensate for the influence of
the bright parts on the dark parts of the image.
This option is always on (and grayed) if a DICOM display function is
selected.
When finished, click on Next.
10 White and black luminance
You can choose to specify one particular white luminance and black
luminance value for all the displays.
Select Yes only if all displays are the same type and if you are sure about
the value to specify.
If you proceed in that case, the Define Luminance page appears.