First installation 54
White target setup on grayscale displays
White target setup on color displays (screenshot taken from Barco Color Coronis displays)
Measure on next calibration
Select this option if you do not wish to enter a target value now. Instead,
the target value will be determined by measuring during the next
calibration.
Absolute value
Select this option if you wish to enter a numeric target value.
This option is preferred for Barco displays.
Luminance
In this field, enter the required target value.
The field will be grayed if the “Measure on next calibration” option is
selected.
Maximized luminance (Barco Color Coronis displays only)
Select this option if you wish to obtain the highest possible luminance. As
a result, the display will always be calibrated to the maximum luminance