Using MediCal Pro 66
3 Using MediCal Pro
3.1 Starting up
To start MediCal Pro, proceed as follows:
1 If the viewing station contains Barco displays, make sure there is no
on-screen display (OSD) on the screen of the displays. Please refer to
the display’s user manual to know how to do this.
2 Start MediCal Pro from the Start > Programs > MediCal menu in
Windows.
3 During start-up, the MediCal Pro Startup application may display the
following message:
This means that another application has changed an imaging board’s
Look-Up Tables since you have last used MediCal Pro.
Select Yes if you want to restore the gamma to the value defined in
MediCal Pro. This is the preferred choice if you were satisfied with the
appearance of the images on the displays when you last closed
MediCal Pro.
Select No if you do not want to restore the gamma defined in MediCal
Pro. In that case, however, your display system will be no longer
calibrated, because it was calibrated according to the gamma defined
in MediCal Pro.
4 During start-up, MediCal Pro may notice a change in the configuration
(e.g., it cannot establish communication with one of the Barco
displays in the configuration).
Here is an example of a message that may appear when MediCal Pro
has detected additional displays in the system:
Select Cancel to stop MediCal Pro.
Select Add now to continue with the start-up.
Here is an example of a message that may appear when MediCal Pro
has detected a display has been removed from the system: