Welch Allyn Medical Diagnostic Equipment 300 Series Pacemaker User Manual


 
Service manual Disassembly procedure 67
...and the flanged side down.
Newer main boards (versions 031-0150-03 and -04)
Disassemble and remove the NIBP assembly
1. Remove the tubing and the pump.
a. Disconnect the tubing from the manifold on the NIBP board.
b. Disconnect the two sections of tubing from the pump.
c. Disconnect the pump power from the NIBP board.
J3 cable connector properly aligned with the
J3 connector on the main board. The cable
connector is slotted on the top side.
J3 cableconnector properly aligned withthe J3
connector on the main board. The cable
connector has latching tabs on the bottom side.
The power connector is keyed to
assure proper connection.