50 Troubleshooting and repair Welch Allyn VSM 300 Series
Monitor HW0Not updatableCompatible
------------------------------------------------------------------------
UPDATE LOG
------------------------------------------------------------------------
March 15 2007 11:32:33 Monitor Main:1.20.00 Success
March 15 2007 11:32:34 cmdSetVSM3ValidVersionPoem:03.20.00 Success
March 15 2007 11:32:34 cmdSetVSM3ValidVersionSpo2:1.9.0.1 Success
March 15 2007 11:32:34 cmdSetVSM3ValidVersionTemp:1.2 Success
------------------------------------------------------------------------
PERSON PERFORMING UPGRADE
------------------------------------------------------------------------
Printed Name ______________________________
Signature ______________________________
Updating NIBP, Temperature & SpO
2
Modules
When replacing the NIBP (POEM), temperature or SpO
2
hardware modules with modules
containing a different firmware version, the service utility must be run to enable the
monitor to accept the new firmware version. If the firmware has been changed since the
last time the monitor was configured, the new firmware version is shown in orange in the
Installed Version column under Update Monitor.
In this screen shot a new temperature module has been installed with firmware
version 1.2. Notice the Upgrade Status displays “Update required”. Until the monitor is
updated, it displays an E35 error indicative of a faulty temperature module.