64 Disassembly procedure Welch Allyn VSM 300 Series
Remove the main board
1. Disconnect the temperature module flex cable from J12 on the main board.
2. Remove the four screws securing the main board to the rear chassis.
Temperature module flex cable
Main boardJ12 connector
Unlatch the ZIF connector before
removing or inserting the flex
cable