32 Functional verification Welch Allyn VSM 300 Series
Here is an example of verification results for a typical monitor.
Test Result
Pass
(or N/A)
Fail
Batterycharging charge symbol lit
Power-onself-test normal
Beeper audible
Monitor-offcurrent (mA): 0
Init/idlecurrent (mA): 467
Baselinecurrent (mA): 199
Batteryvoltage Monitor (V): 5.94 DMM (V): 5.95
NIBPcharacterization test C03
NIBPverification
Pressureleakage Before (mmHg): 100 After (mmHg): 99
Pressureaccuracy 80 mmHg SYS:73 Manometer: 72
150 mmHg SYS:142 Manometer:140
300 mmHg SYS:290 Manometer: 286
Valve/pumpcurrent Valve/Pump (mA)
Baseline (mA)
Difference (mA)
560
199
361
Inflation Elapsed time (sec): 6
Deflation Manometer reading (mmHg): 0 (3 sec)
Maxallowable cuff pressure
Adult (280 [37.3]) 280 mmHg kPa
Pediatric (200 [26.7]) 200 mmHg kPa
Neonate (141 [18.8]) 141 mmHg kPa
Overpressurecut-off (Pass-Fail)
Printercurrent Peak (mA): 940
SpO
2
Saturation (%): 90 Pulse (bpm): 200
Temperature Fahrenheit: 105.9 Celsius:
Nurse call relay pins1-2 Alarm() >1M NoAlarm () <1
pins 2-3 Alarm () <1 NoAlarm () >1M
Batterycharge time 12.5 h
Batterydischarge time 19 h
Batteryvoltage 6.40 volts
Monitor Serial #: JA036582
Tested by: RJ
Test date: 8/30/2007