4. Turn the potentiometer fully clockwise until it stops, and then back 1/8 turn
counter-clockwise.
5. Tighten the set screw to secure the recline linkage to the recline potentiometer.
6. Secure the recline linkage to the center seat frame using the socket screw, spacer and
locknut.
7. Route and secure the recline potentiometer cable as described in Securing the Cables
on page 157.
8. Plug the recline potentiometer cable into the TRCM or TAC (3-pin Molex connector).
9. Adjust the recline potentiometer. Adjusting Recline Potentiometer on page 154.
10. Perform the post-service inspection. Refer to Post-Service Inspection Checklist on
page 21.