Invacare 2GTR Wheelchair User Manual


 
SECTION 5—INTERFACE HARDWARE
2G Tarsys®Powered Seating System 34 Part No 1114842
*NOTE: Shims are not present on seating systems manufactured after 12/18/2005.
CAUTION
The linear guides MUST be parallel and the rear pivot assembly MUST move freely
on the linear guides. Otherwise, damage to the seating system will occur during
operation.
11. Adjust the linear guides until they are parallel.
12. Slide the rear pivot assembly forward and back to see if it slides freely on the linear
guides.
13. Perform one of the following:
Rear Pivot Assembly Slides Freely - Proceed to STEP 14.
Rear Pivot Assembly Is Hard To Move - Repeat STEP 11 and STEP 12 until the rear
pivot assembly moves freely.
NOTE: It may be necessary to loosen the screws installed in STEP 10 to adjust the linear guides.
14. Torque the socket screws to 75 in-lbs ± 20%.
15. Install the shoulder bolt and locknut to secure the front of the tilt actuator to the front
pivot assembly. Torque to 23 ft-lbs ± 20%.
16. Perform the post-service inspection checklist. Refer to Post-Service Inspection
Checklist on page 21.
FIGURE 5.1 Repositioning the Seat for 2GT and 2GTR Systems - Securing the Linear Guides
Lower Portion of Linear Guide
Rear Pivot Assembly
Socket Screw
Access Hole
Mounting Hole