Invacare 2GTR Wheelchair User Manual


 
SECTION 6—SEAT ADJUSTMENTS
Part No 1114842 75 2G Tarsys®Powered Seating System
Removing/Installing the Seat
NOTE: If retrofitting a seating system, DO NOT use this procedure. Refer to the 2G Tarsys
Seating System Installation on TDX instruction sheet, part number 1134889.
Removing
Disconnecting Cables
NOTE: For this procedure, refer to FIGURE 6.17 on page 76.
NOTE: Invacare recommends two people perform this procedure.
1. Tilt Only and Tilt/Recline Systems - Tilt the seat back 10° to 15° to allow easier access
to the hardware securing the seating system to the base.
2. Turn off the joystick.
3. Disconnect the joystick.
4. Disconnect the TRCM, TAC or SAC from the MK
5
controller (5-pin connector).
5. Disconnect the auxiliary power connectors from the power take off connectors (not
shown) on the controller.
6. Perform one of the following:
Tilt Only and Tilt/Recline Systems -
i. Disconnect the tilt actuator connector (2-pin Anderson) from the TRCM, TAC,
SAC or TRSS.
ii. Disconnect the tilt potentiometer connector (3-pin Molex) from the TRCM,
TAC, SAC or TRSS.
Recline Only Systems - Disconnect the recline potentiometer connector (3-pin
Molex) from the TRCM or TAC.