SECTION 5—INTERFACE HARDWARE
2G Tarsys®Powered Seating System 24 Part No 1114842
SECTION 5—INTERFACE
HARDWARE
ƽ WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
Before adjusting, repairing or servicing the seating system, ALWAYS turn the
wheelchair power OFF, otherwise, injury or damage may result.
Pinch points exist between seat and base frames. Use caution, otherwise injury may
occur.
Repositioning the Seat
About Repositioning the Seat
The seat position MUST be adjusted under the following conditions:
The seat depth changes.
The seating system changes so it has both a ventilator-ready battery tray and center
mount footrests.
The mounting location of the seat depends on the seat depth of the seating system and
whether or not both a ventilator-ready battery tray and center mount footrests are
installed. The ventilator battery tray requires a long battery box door. When used in
combination with center mount riggings, the seating system MUST be positioned further
forward for proper clearance.
Repositioning the Seat on 2GR Systems
NOTE: For this procedure, refer to Table 1 - Table 7 on page 26 to page 32.
1. Refer to the tables at the end of this procedure to determine if a seat position
adjustment is needed.
NOTE: The seat adjustment tables are based on the seat depth of the system. Some seat depths share
the same mounting position.
2. Remove the six hex screws, washers and locknuts securing the rear pivot assembly,
spacer bars and seat brackets to the lower interface brackets.
3. Remove the two hex screws and locknuts securing the insides of the seat brackets to
the lower interface brackets.
4. If necessary, remove the four hex screws and locknuts securing the front pivot
assembly to the seat brackets and lower interface brackets.