Invacare 2GTR Wheelchair User Manual


 
SECTION 14—ELECTRONICS
Part No 1114842 155 2G Tarsys®Powered Seating System
FIGURE 14.8 Adjusting Recline Potentiometer
Removing/Installing the Recline Potentiometer
NOTE: For this procedure, refer to FIGURE 14.9 on page 156.
Removing
1. Remove the socket screw, spacer and locknut securing the linkage to the center seat
frame.
2. Loosen the set screw on the recline linkage.
3. Remove the recline linkage from the recline potentiometer.
4. Remove the nut and washer securing the recline potentiometer to the bracket.
5. Remove the recline potentiometer from the bracket.
6. Unplug the recline potentiometer cable from the TRCM or TAC.
7. Cut any tie-wraps securing the recline potentiometer cable to the wheelchair.
Installing
1. Install the recline potentiometer up through the bracket.
2. Install the washer onto the recline potentiometer and secure to the bracket using the
nut.
3. Install the recline linkage onto the recline potentiometer.
DETAIL “A”
Set Screw
Recline
Potentiometer