Invacare 2GTR Wheelchair User Manual


 
SECTION 7—BACK ADJUSTMENTS
2G Tarsys®Powered Seating System 82 Part No 1114842
Removing/Installing the Curved Links
Tilt Only Systems
NOTE: For this procedure, refer to FIGURE 7.2 on page 83.
Removing
ƽ WARNING
Perform this procedure on one side of the wheelchair at a time. Otherwise the back
will be completely disconnected from the seat and injury or damage may occur.
1. Remove the side shroud. Refer to Removing/Installing Side Shrouds on page 100.
2. Remove the small socket screw and locknut securing the top of the curved link to the
back cane.
3. Remove the large socket screw, nylon washer and locknut securing the rear of the slide
block to the connecting bracket.
4. Remove the curved link from the wheelchair.
Installing
1. Secure the top of the curved link to the back cane using the small socket screw and
locknut. Torque to 23 ft-lbs ± 20%.
2. Secure the bottom of the curved link to the rear of the slide block and the connecting
bracket using the large socket screw, nylon washer and locknut in the same position
used on the opposite side of the wheelchair. Torque to 156 in-lbs ± 20%.
ƽ WARNING
Side shroud(s) MUST be reinstalled before operation of the seating system. Other-
wise injury may result.
3. Install the side shroud. Refer to Removing/Installing Side Shrouds on page 100.