HP (Hewlett-Packard) RM500SL Hearing Aid User Manual


 
0611 RM500SL Users Guide Version 2.8 Page 86
19.4 Microphone location effects
FastFacts 19.4: Microphone location effects
Microphone Location Effects
Add to all stimuli in S-REM mode
Hearing instrument type
Frequency
BTE ITE ITC CIC BODY
200 0.5 0.3 0.0 0.9 3.0
250 0.5 0.5 0.3 0.8 3.0
315 0.8 0.8 0.3 0.7 3.0
400 1.1 1.0 0.7 0.5 3.0
500 1.2 1.8 0.0 0.4 4.0
630 1.1 2.0 0.1 0.4 3.0
800 0.9 2.0 0.4 0.4 2.0
1000 0.3 1.5 1.2 0.6 0.0
1250 0.6 0.3 -1.6 1.1 0.0
1600 2.5 -0.3 -1.9 2.0 -4.0
2000 4.1 3.8 2.1 3.3 -4.0
2500 3.5 5.0 4.8 5.0 -3.0
3150 2.8 3.3 3.5 6.9 -2.0
4000 3.7 4.3 6.4 8.3 0.0
5000 -1.2 4.3 6.6 7.6 0.0
6300 1.6 -0.4 -1.8 4.2 0.0
8000 3.3 1.0 -1.9 -5.6 0.0
Microphone location effects are added after the overall rms level has been set