3.6 Screen messages
FastFacts 3.6: Screen messages
3.7 Program memory removal and replacement
FastFacts 3.7: Program memory removal and replacement
The
Title ba
r
(top line of the display screen) informs you of the selected Test (e.g.,
Multicurve).
The Message bar (bottom line of the display screen) suggests the next step in a
test or informs you of the state of the instrument or conditions that could affect your
data. For example, it will inform you if the microphones need to be calibrated.
Operating software is contained on a USB flash drive located in the test chamber
under the printer paper roll cradle as shown. Software updates are shipped on a
CD-ROM and must be transferred to the USB flash drive in order to be used. This
requires that the USB flash drive be removed from the RM500SL and plugged into
the USB port on a PC that has a CD-ROM drive.
To remove the USB flash drive, ensure that the RM500SL power is off and:
1) Open the test chamber and remove the foam from the test chamber floor.
2) Pull the USB flash drive straight toward the front of the unit until it is free.
To install the USB flash drive, ensure that the RM500SL power is off and:
1) Open the test chamber and remove the foam from the test chamber floor.
2) Gently insert the USB flash drive into the USB port, with its foam pad down.
3) Replace the foam in the floor of the test chamber.
NOTE that the USB
flash drive has been
especially configured
for the RM500SL.
Replacing this flash
drive with another
can cause erratic
operation. Such
replacement should
be done only when
authorized and
ONLY by an
Audioscan service
representative.