HP (Hewlett-Packard) RM500SL Hearing Aid User Manual


 
0611 RM500SL Users Guide Version 2.8 Page 69
16.7 SII calculation in Insertion gain
FastFacts 16.7: SII calculation in Insertion gain
The RM500SL calculates a Speech Intelligibility Index (SII) value (in %) for
unaided speech at 70 dB SPL (REUR), the target REIG and for each aided test.
The calculations use the 1/3 octave band method of ANSI S3.5 1997 without the
160 Hz band and without masking effects. Level distortion effects are included but
no hearing loss desensitization is applied. An idealized speech region of LTASS
+/- 15 dB is used.
The speech recognition associated with a given SII is a function of the test
material and the cognitive abilities of the listener. There is considerable individual
variability in relating the SII to speech recognition, especially amongst impaired
listeners. The following shows expected nominal recognition scores vs SII for
normals on the Connected Speech Test (Sherbecoe and Studebaker 2003).
SII (%)
Score (%)
10 1
15 8
20 20
25 45
30 60
35 74
40 84
45 90
50 94
55 96
60 98
70 99
80 100
90 100