Emerson Process Management OCX 8800 Oxygen Equipment User Manual


 
Instruction Manual
IM-106-880C, OI
September 2009
4-5
Hazardous Area OCX 8800
Figure 4-4. LOI Menu Tree
(Sheet 2 of 4)
NOTE: “Hit E when ready”is displayed during
semi-automatic calibration only (when
Calib Setup value Use Solenoids = n).
(CONTINUED FROM
SHEET 1)
CALIBRATION
(CONTINUED ON
SHEET 3)
Cal
Control
Cal
Constants
Cal
Status
Cal
Verify
Abort Cal
Start Cal
O2
Start Cal
COe
Start Cal
Both
Current
Calib
Prev
Calib
Falied
Calib
Calib Result
Calib Step
Calib Time
Next O2 Cal
Next COe Cal
Flow High Gas
Flow Low Gas
Flow COe Gas
Purge
Status
Apply Lo O2 Gas
Flow O2 Lo _____s
______% _____mV
Read O2 Lo _____s
______% _____mV
Apply Hi O2 Gas
Flow O2 Hi _____s
______% _____mV
Read O2 Hi _____s
______% _____mV
Purge _____s
Hit E when ready
Hit E when ready
Hit E when ready
Apply Lo COe Gas
Flow COe Lo _____s
______% _____mV
Read COe Lo _____s
______% _____mV
Hit E when ready
Apply Hi COe Gas
Flow COe Hi _____s
______% _____mV
Read COe Hi _____s
______% _____mV
Purge _____s
Hit E when ready
Hit E when ready
Apply Lo O2 Gas
Flow O2 Lo _____s
______% _____mV
Read O2 Lo _____s
______% _____mV
Apply Hi O2 Gas
Flow O2 Hi _____s
______% _____mV
Read O2 Hi _____s
______% _____mV
Hit E when ready
Hit E when ready
Purge _____s
Flow COe Hi _____s
______% _____mV
Read COe Hi _____s
______% _____mV
Purge _____s
Hit E when ready
Hit E when ready
O2 Slope _____mV/D
O2 Constant _____mV
O2 Sensor R _____ohm
COe Slope _____ppm/mV
COe Constant _____mV
Pre O2 Slope _____mV/D
Pre O2 Const _____mV
Pre O2 Sensor R _____ohm
Pre COe Slope _____ppm/mV
Pre COe Const _____mV
Bad O2 Slope _____mV/D
Bad O2 Const _____mV
Bad COe Slope _____ppm/mV
Bad COe Const _____mV