Lifescan Blood Glucose Monitor Blood Glucose Meter User Manual


 
Troubleshooting
68
Chapter 5
LifeScan Healthcare Professional Line: 1 800 524-7226
The meter detected
a temperature
problem.
1.Repeat test with new
test strip. Make sure
meter and test strip are
at room temperature.
• Do not apply heat or
cold to meter. Operating
temperature is 18°–30°C.
2.If you continue to get
any error message,
obtain a laboratory
result, and
• Compare color of test
strip confirmation dot to
Color Chart on test strip
bottle to see if glucose
level is very high or very
low.
• Take appropriate steps
to monitor patient and
provide medical care
while awaiting
laboratory result.
3.Contact LifeScan for help.
Symbol Possible Cause What To Do
SSHosp.book Page 68 Thursday, November 14, 2002 3:44 PM