Operation
28
Chapter 2
Control Solution Results
When the control test result falls within the
control range, the meter has passed the QC
check and can be used to test patient blood.
If the result falls outside the expected range,
repeat the test. Results that fall outside the
expected range may indicate:
procedural error
control solution is expired, contaminated,
beyond its discard date, or inadequately
mixed
code number in meter does not match
code number on test strip bottle
dirty meter or test strip holder
problem with test strip (expired,
damaged, etc)
control solution testing temperature
outside the range of 18°–30°C (64°–86°F)
too much or too little control solution
applied to test strip