Lifescan Blood Glucose Monitor Blood Glucose Meter User Manual


 
Operation
30
Chapter 2
Run a linearity solution test as you would a
control solution test. Follow the instructions
under “Performing a Quality Control Test”
on page 23, testing the linearity levels in
replicates as specified by your institution’s
policy.
Compare the linearity results with the
linearity target values printed on the
SureStepPro Test Strip bottle, or with your
institution’s established ranges. Linearity
solution results can be affected at altitudes
of 2000 feet or greater. Read the
SureStepPro Linearity Test Kit package
insert for more information.
Patient Blood Test
LifeScan recommends that only certified
operators perform blood glucose testing
using the SureStep Hospital Meter. Follow
your institution’s policies for in-service
training to become a certified operator.
SSHosp.book Page 30 Thursday, November 14, 2002 3:44 PM