Lifescan blood glucose meter Blood Glucose Meter User Manual


 
DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
85
Why is insulin delivered when the
dose selector is turned?
You have dialed the maximum dose
of 70 units and then continued
dialing. If you need to inject a dose
of insulin larger than 70 units, you
must divide the dose into amounts
equal to, or less than, 70 units.
Why does the push button block
during the injection?
Do not try to force the push button
in. The needle may be blocked.
Change the needle and prime the
InDuo
TM
doser as described on page
42. Check to see if PenFill
®
is
damaged. Dial back to zero 0 and
press the push button completely in
until it locks. Then open the slide
and change PenFill
®
as described on
page 49. Prime the InDuo
TM
insulin
doser until insulin appears at the
needle tip (see page 24). Please note
that part of the insulin dose may
have been injected. If the push
button still cannot be pressed
completely in, call InDuo
TM
Customer
Services at 1-877-520-9056.
Why does the slide not open when
changing PenFill
®
?
Put on a new needle. Press the push
button completely in until it locks.
Remove the needle and open the
slide. If the slide still cannot be
opened, call InDuo
TM
Customer
Services at 1-877-520-9056.
Do not try to force the slide open.
DCO_06084101C_InDuo_OB_US 8/16/04 1:08 PM Page 85