Lifescan blood glucose meter Blood Glucose Meter User Manual


 
TESTING YOUR BLOOD GLUCOSE LEVELS 35
h
i
Allow Blood Drop to Form
Continue holding the sampler against the skin for a
few seconds until the blood drop forms. Allow
enough blood to form under the cap until you have
a blood sample that is sufficient to fill the
confirmation window of the test strip. If bruising
occurs, you may choose to lance a fingertip instead.
If you are having difficulty obtaining blood from the
arm, call InDuo
TM
Customer Services at 1-877-520-9056
for assistance. The blood sample must be at least
1 microliter in volume or you may get an Æ or
possible inaccurate test result.
Removing the lancet
To remove the cap, twist it counterclockwise h.
Optional: To replace the protective disk on the
lancet, place the disk on a hard surface and push
the exposed tip into the protective disk.
æAlways use caution when removing the lancet.
Point the lancet down and away. Push forward on
the ejection/cocking control and eject the lancet
directly into an appropriate container for sharp
objects i. Return the ejection/cocking control to
the middle position. Replace the cap.
DCO_06084101C_InDuo_OB_US 8/16/04 1:08 PM Page 35