DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
Meter display What it means Action
Error message that indicates that
the InDuo
TM
meter has detected a
problem with the test. Possible
causes are incompletely filled
confirmation window or a
damaged OneTouch
®
Ultra test
strip.
Repeat the test with a new
OneTouch
®
Ultra test strip.
Refer to page 34 and 37 for
proper sample application
procedure. If the error
message appears again, call
InDuo
TM
Customer Services at
1-877-520-9056.
The battery symbol É appears on
the display with the unit of measure.
It means that the power of the
battery is getting low. You can
complete about 50 more tests from
the time this symbol first appears.
Test results will still be
accurate, but replace the
battery as soon as possible.
The battery symbol É appears
on the display by itself. The power
of the battery is too low to run a
test.
Replace the battery
immediately. The Induo
TM
meter will not operate until
the battery is replaced.