Lifescan blood glucose meter Blood Glucose Meter User Manual


 
GETTING YOUR InDuo
TM
SYSTEM READY FOR USE 15
æ Caution
Use only OneTouch
®
Ultra Control Solution
Check the expiration date on the control solution
vial. Do not use if expired.
Store control solution tightly closed at
temperatures below 86ºF (30°C).
Do not refrigerate control solution.
The control solution, the InDuo
TM
meter, and the
OneTouch
®
Ultra test strips should be at room
temperature 68–77ºF (20–25ºC) before testing.
Shake the vial, discard the first drop of control
solution, and wipe off the dispenser tip with a clean
tissue to obtain a good sample and an accurate result.
Use only for 3 months after the first opening.
Record the discard date (date opened plus
3 months) on the control solution vial. Discard
after 3 months.
The control solution range printed on the
OneTouch
®
Ultra test strip vial is for OneTouch
®
Ultra control solution only. It is used to test meter
and OneTouch
®
Ultra test strip performance. It is
not a recommended range for your blood glucose
level.
DCO_06084101C_InDuo_OB_US 8/16/04 1:08 PM Page 15