Invacare Spree XT Mobility Aid User Manual


 
SECTION 10—TILT
Spree XT 68 Part No 1118364
Adjusting the Limit Stops
ƽ WARNING
Contact a HEALTHCARE PROFESSIONAL or ASSISTANT for allowable tilt angles
that are suitable for the occupant.
TRANSPORT READY OPTION ONLY - Wheelchairs with adjustable seat angles
MUST be set between 0° and 15°.
NOTE:Forthisprocedure,refertoFIGURE10.3.
1. Engagebothwheellocks.
2. Tiltwheelchairtodesiredposition.
3. Loosenthesetscrewsthatsecurethe
limitstopinplace.
4. Positionlimitstopflushagainsttopof
cylinder(s).
5. Retightenthesetscrews.Securely
tighten.
6. Ifapplicable,repeatforothertilt
mechanism.
FIGURE 10.3 Adjusting the Limit Stops
Limit Stop
Set
Screw