Invacare Power Tiger Mobility Aid User Manual


 
SECTION 7—FRONT RIGGINGS
Power Tiger™ 54 Part No 1134839
Installing/Adjusting Adjustable Angle Flip-Up Footplates
Installing Flip-Up Footplates
NOTE: For this procedure, refer to FIGURE 7.2.
1. Slide the half clamp over the footplate hinge.
2. Loosely tighten the two flat screws that secure the footplate to the half clamp.
3. Adjust the footplates to the necessary angle and depth for the user. Refer to Adjusting
Flip-Up Footplate Angle on page 54.
Adjusting Flip-Up Footplate Depth
NOTE: For this procedure, refer to FIGURE 7.2.
1. Remove the two flat screws, washers
and locknuts that secure articulating
footplate to the half clamp.
NOTE: Note the angle of the articulating
footplate for reinstallation.
2. Move articulating footplate to one of
four mounting positions.
NOTE: If desired depth is still not obtained,
rotate the half clamp on the footplate hinge
180°.
3. Retighten the two flat screws, washers
and locknuts.
FIGURE 7.2 Adjusting Flip-Up Footplate
Depth
NOTE: The settings for positioning the articulating footplates on the half-clamps may vary for
each footplate.
4. Adjust footrest. Refer to Adjusting Flip-Up Footplate Angle on page 54 or Adjusting
Flip-Up Footplate Perpendicular and/or Inversion/Eversion on page 55.
Adjusting Flip-Up Footplate Angle
NOTE: For this procedure, refer to FIGURE 7.3 on page 55.
NOTE: It is not necessary to remove the footplate to perform this adjustment.
1. Insert a flathead screwdriver through the half clamp on the articulating footplate.
2. Slowly turn nylon adjustment screw in or out until articulating footplate is
perpendicular to the footrest assembly or the desired inversion or eversion is
obtained.
Washers
Locknuts
Half Clamp
Footplate Hinge
Nylon
Adjustment
Screw
Footplate Hinge
Articulating
Footplate
Flat Screws
90° Footrest Support