Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15
Hazardous Area Oxymitter DR
f. WORLD CLASS 3000 INTELLIGENT
FIELD TRANSMITTER SETUP
The DR probe operates with a 115 VAC
heater. Ensure that the voltage selection
jumpers in the IFT or HPS, if used, are set
properly. Refer to Figure 2-13 for IFT
jumper selection, and Figure 2-14 for HPS
jumper selection. For additional setup and
configuration information, refer to IB-106-
300NFX.
ALWAYS DISCONNECT LINE VOLTAGE
FROM INTELLIGENT FIELD TRANSMITTER
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATION
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
PROBE HEATER
VOLTAGE SELECTION
JUMPER
(INSTALL)
100 V.A.C.
120 V.A.C.
220 V.A.C.
240 V.A.C.
JM3, JM7, JM2
JM8, JM7, JM1
JM6, JM5, JM2
JM6, JM5, JM1
WORLD CLASS PROBE (44V)
WORLD CLASS "DIRECT
REPLACEMENT" PROBE (115V)
OR OXYMITTER DIRECT
REPLACEMENT PROBE (115V)
218 PROBE (115V)
JM10
JM9
JM9
36210015
Figure 2-13. IFT 3000 Power Supply Board Jumper Configuration
If incorrect heater voltage is selected, damage to the Oxymitter DR may occur. Refer to
IB-106-300NFX for additional configuration information. Always update the relevant labeling
to reflect the set voltage.
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY AND
ANALOG ELECTRONICS (IF USED)
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATIONS
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
JUMPER
100/120 V.A.C.
220/240 V.A.C.
JM4, JM1
JM5
REMOTE
ON
REMOVE JM2
*INSTALL JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SELECTION
INSTALL JM3, JM6
REMOVE JM3, JM6
*ANALOG (EXISTING)
DIGITAL
(NEXT GENERATION)
WORLD CLASS PROBE
(44V)
JM7
JM8
218 PROBE (115V)
1
NOTES:
100 V.A.C. OPERATION REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.
1
36210014
JM8
DIRECT REPLACEMENT
WORLD CLASS OR
DR OXYMITTER
Figure 2-14. Heater Power Supply (HPS 3000) Jumper Configuration
2