Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Hazardous Area Oxymitter DR
TO ELECTRONICS
REF AIR SET
263C152G01
1 FLOWMETER 0.2-2.0 SCFH 771B635H02
2 2" PRESSURE GAGE 0-15 PSIG 275431-006
3 COMBINATION FILTER-REG. 0-30 PSIG 4505C21G01
NOTE: DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
1
2
3
4.81 (122.17)
FLOW SET
POINT KNOB
0.125-27 NPT FEMALE
OUTLET CONNECTION
1.19
(30.22)
10.0
(254)
REF
DRAIN VALVE
3.12 (79.25) MAX
8.50
(215.90)
MAX
2.0
(50.80)
2 MOUNTING HOLES
3.19 (81.03) LG
THROUGH BODY FOR
0.312 (7.92) DIA BOLTS
1.50
(38.10)
2.250 (57.15)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER DR PROBE HEAD.
OUTLET
0.25-18 NPT FEMALE
INLET CONNECTION
INSTRUMENT AIR SUPPLY
10-225 PSIG MAX PRESSURE
36210008
0.250 OR 6 MM O.D. TUBING
(SUPPLIED BY CUSTOMER)
FITTING
TUBE COMPRESSION
0.250 OR 6 MM O.D.
SERIAL NO.
TAG NO.
OXYMITTER DR
Rosemount Analytical Inc.
Orrville,OH 44667-0901
TM
800-433-6076
Figure 2-8. Air Set, Plant Air Connection
Install all protective equipment covers
and safety ground leads before
equipment startup. Failure to install
covers and ground leads could result
in serious injury or death.
2-4 SYSTEM SETUP
a. Overview
This section covers the setup procedures
for the Oxymitter DR In-Situ Oxygen Probe.
The DR probe can be used with several
electronics packages including: Models 218,
218A, 225, 132 (analog and digital), TC200,
and ZA8C.
For Westinghouse Models 218, 225, and
132 (analog), refer to paragraph 2-4b. Addi-
tional information can be found in IB-106-
101.
For Westinghouse Model 218A, refer to
paragraph 2-4c. Additional information can
be found in IB-106-101A.
For Westinghouse Model TC200, refer to
paragraph 2-4d. Additional information can
be found in IB-107-020.
For Westinghouse Model 132 (digital), refer
to paragraph 2-4e. Additional information
can be found in IB-106-106A.
For Rosemount World Class 3000, refer to
paragraph 2-4f. Additional information can
be found in IB-106-300NFX.
For Yokogawa Model ZA8C, refer to para-
graph 2-4g.
2