Emerson 705 Pacemaker User Manual


 
52
Reference Manual
00809-0200-4705, Rev AA
Appendix B: Product Certifications
May 2015
Product Certifications
IU ATEX Intrinsic Safety for Zone 2
Certificate: Baseefa15ATEX0059X
Standards: EN 60079-0: 2012, EN 60079-11: 2012
Markings: II 3 G Ex ic IIC T4 Gc, T4(-60 °C T
a
+70 °C)
For use with Rosemount SmartPower power module part number 753-9220-0001, or for
use with Emerson SmartPower option 701PBKKF.
Special Conditions for Safe Use (X):
1. The surface resistivity of the antenna is greater than 1GΩ. To avoid electrostatic
charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
2. The 705 enclosure may be made of aluminum alloy and given a protective
polyurethane paint finish; however, care should be taken to protect it from impact or
abrasion if located in a Zone 0 area.
International
I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 14.0173X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4 (-60 °C T
a
+70 °C)
For use with Rosemount SmartPower power module part number 753-9220-0001, or for
use with Emerson SmartPower option 701PBKKF.
Special Conditions for Safe Use (X):
1. The surface resistivity of the antenna is greater than 1GΩ. To avoid electrostatic
charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
2. The 705 enclosure may be made of aluminum alloy and given a protective
polyurethane paint finish; however, care should be taken to protect it from impact or
abrasion if located in a Zone 0 area.
Totalizer terminal parameters Turbine meter parameters
U
O
= 2.5 V U
i
= 10 V
I
O
= 5.6 mA I
i
= 10 mA
P
O
= 13.9 mW P
i
= 1 mW
C
i
= 2.2 nF N/A
L
i
= 4.7 mH N/A
Turbine meter terminal output
parameters
Turbine meter terminal input
parameters
U
O
= 2.5 V U
i
= 10 V
I
O
= 5.6 mA I
i
= 10 mA
P
O
= 13.9 mW P
i
= 1 mW
C
i
= 2.2 nF N/A
L
i
= 4.7 mH N/A