Emerson 705 Pacemaker User Manual


 
24
Reference Manual
00809-0200-4705, Rev AA
Section 4: Commissioning
May 2015
Commissioning
Figure 4-4. Network Join Status Screens
Connecting with a Field Communicator
A 705 DD is required for HART
®
communication.
Checking for communication using the Smart Wireless Gateway
In the integrated web interface from the Gateway, navigate to the devices page. This page
shows whether the device has joined the network and if it is communicating properly.
Note
The time to join the new device(s) to the network is dependent upon the number of devices
being joined and the number of devices in the current network. For one device joining an
existing network with multiple devices, it may take up to five minutes. It may take up to 60
minutes for multiple new devices to join the existing network.
Searching for
network
Joining network
Connected with
limited bandwidth
Connected
Function Key sequence Menu items
Communications
2, 1, 3 Join Status, Communication Status, Join Mode, Number of
Available Neighbors, Number of Advertisements Heard,
Number of Join Attempts
N E T w K
S R C H N G
n e t w k
N E G O T
n e t w k
L I M - O P
n e t w k
O K