Section 2: Configuration
May 2015
Configuration
If a custom volume unit is desired, the 705 totalizer can be configured to report the totalized
Pulses, and a custom conversion can be done on the host system side. Similarly, if a custom flow
rate unit is desired, the 705 totalizer can be configured to report pulses per second (“Hertz”),
and a custom conversion can be done on the host system side. When reporting Pulses or Hertz,
the configured K-factor has no affect on the reported values.
2.4 Joining the device to a network
To communicate with the Gateway, and ultimately the Host System, the transmitter must be
configured to communicate with the wireless network.
Using a Field Communicator or AMS Wireless Configurator, enter the Network ID and Join Key so
they match the Network ID and Join Key of the Gateway and the other devices in the network. If
the Network ID and Join Key are not identical, the transmitter will not communicate with the
network. The Network ID and Join Key may be obtained from the Gateway on the Network>Set-
tings page on the web server or by clicking network information in the upper right hand corner.
Using a Field Communicator, the Network ID and Join Key can be configured using a Field
Communicator with the Fast Key Sequence: 2, 1, 3.
The final device network configuration piece is the Update Rate which, by default, is 1 minute. It
can be changed at commissioning, or at any time, by using AMS Wireless Configurator or the
Gateway’s web server. The Update Rate should be between 1 second and 60 minutes. To change
the Update Rate with a Field Communicator, use the Fast Key Sequence: 2, 1, 4.
If doing a bench top initial configuration, after completion remove the power module until
installation. When the device is installed, insert the power module and close the housing cover
securely. Always ensure a proper seal so that metal touches metal, but do not over tighten.
Configure units fast keys
Volume units fast keys
2, 1, 2
Flow units fast keys
2, 2, 1