Accu-Chek insulin pump Blood Glucose Meter User Manual


 
You can cancel the extended delivery by putting the pump in STOP
mode. This cancels the Extended Bolus. An alert A8: BOLUS CANCELLED
occurs. Press
f twice to confirm and turn off the alert.
Bolus delivery is interrupted.
See the section “Alert A8: BOLUS CANCELLED” (page 153) for
more information.
You can review the actual bolus amount delivered before the cancellation
by looking at the bolus history. See the section “Review the Bolus History”
(page 107) for more information.
Make sure that the cancellation was intended and then program a new MultiWave
Bolus (and/or Temporary Basal Rate), if needed.
Note:
If a Temporary Basal Rate is active at the same time, it is also cancelled
when your pump is put in STOP mode. An alert A8: BOLUS CANCELLED
and an alert A6: TBR CANCELLED occur.
Press
f twice to confirm and turn off the first alert. The second alert
appears on the display. Press f twice to confirm and turn off the second
alert. Both alerts are recorded in the alarm history.
See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:
BOLUS CANCELLED”
(pa
ge 153) for more information.
!
When your insulin pump is in STOP mode, it does not deliver
an
y insulin. Put your insulin pump in R
UN mode to continue
the insulin deliver
y.
Note:
If the alert A8: BOLUS CANCELLED does not occur, the bolus delivery
was not cancelled. Cancel the extended delivery (see below).
Chapter 4: Boluses and TBRs ACCU-CHEK Spirit Insulin Pump92