Chapter 7: Alerts and Errors ACCU-CHEK Spirit Insulin Pump158
If the error E4: OCCLUSION occurs again, prime your pump without a
cartridge. Follow the steps below:
1.
Press f twice to turn off and to confirm the error.
2. Disconnect or remove the infusion set from your infusion site.
3. Remove the cartridge, adapter and infusion set from your pump.
4. Carry out the PRIME THE INFUSION SET function without the cartridge,
adapter, and infusion set inserted.
5. Insert a new cartridge with connected adapter and infusion set.
6. Prime the infusion set.
7. Reconnect the infusion set (tubing) at your infusion site.
8. Put your pump in RUN mode.
If the error E4: OCCLUSION continues to occur, contact your doctor or
healthcare team for an alternate therapy plan. Contact Pump Support at
1-800-688-4578 for more assistance.
Error E5: END OF OPERATION
Your pump timer has reached zero. The pump will change to STOP mode and will
no longer operate. An error E5: END OF OPERATION will appear on the display. It
is recommended that you replace your pump after four years. Before this error
occurs, an alert A5: PUMP TIMER will remind you to replace your pump soon.
1. Press f twice to turn off and to confirm the error.