Battery
Your pump runs on one AA battery. Use AA alkaline batteries that have a minimum
capacity of 2500 mAh. Do not use lithium, carbon zinc, or nickel cadmium (NiCd)
batteries. If you prefer to use rechargeable batteries, use NiMH batteries that
ha
ve a minimum capacity of
1500 mAh. If you use rechargeable batteries, always
use a battery charger recommended by the battery manufacturer. Keep extra
batteries on hand.
The operating temperature of the battery must be between 41°F and 104°F
(+5°C and +40°C). The life of any battery is affected by how you use your
insulin pump, your personal settings, delivery rates, temperature and other factors.
With typical pump use [50 U/day using U100 insulin; operating temperature
72°F ± 6°F (22°C ± 3°C)], pump battery life is about four weeks using alkaline
batteries and one week using rechargeable batteries.
!
Using batteries other than those recommended by Disetronic
Medical Systems may significantly reduce battery usage
time.
Batteries other than those recommended could leak
and corrode the battery contacts within your insulin pump.
Therefore, using batteries not supplied or recommended
by Disetronic Medical Systems, Inc. may void the warranty.
Chapter 1: Understanding Your New System ACCU-CHEK Spirit Insulin Pump
21
CAUTION